Overview    I    Features    I     Accessories     I     Downloads     
https://sites.google.com/a/laxcoinc.com/www/alpha-series/alpha-1506/specification/Grainger.png




Key Features:

    • Standalone units with 4 line x 20 characters LCD display
    • Wide range in wavelengths from 190nm to 1100nm
    • Single beam systems, Littrow, grating 1200/line/mm
    • Advanced quantitative analysis with linear and output port for data curve 
    • UV/Vis light spectrophotometer 
    • DNA/Protein, Kinetic (enzyme reaction rate) & Wavelength Mode
    • Compartmented multi cuvette sample holder 
    • Digitally selectable wavelength & intensity adjustment
    • Built in Thermal printer output port for data curve (printer not included)
    • Supports up to 1000 test points that can be saved onto local memory
    • Research grade; perfect for labs and industry
    • Halogen ad deuterium lamp to supply light
    • A monochromator to isolate the wavelength of interest and eliminate the unwanted second order radiation
    • A sample compartment to accommodate the sample solution
    • A detector to receive and convert the transmitted light to an electrical signal
    • A digital display to show absorbance and transmittance

SPECIFICATION

  Wavelength Range
  Spectral Slit Width
  Wavelength Accuracy
  Wavelength Repeatability
  Wavelength Resolution
  Photometric Range    
  
Photometric Accuracy
  Photometric Repeatability

  Stray Light
  Stability
  Data Output  
  
Power Requirements
  
Dimensions
  
Net Weight
  190 ~ 1100 nm
  4 nm
  
± 0.5 nm
  
≤ 0.2 nm
  0.1 nm
  0-200%T, -0.3-3A, -9999-9999C
  ±0.002A(0~0.5A), ±0.004A(0.5~1A)
  ±0.008A(1~2A), ±0.3%T(0~100%T)
  ±0.001A(0~0.5A), ±0.002A(0.5~1A)
  ±0.004A(1~2A), ±0.15%T(0~100%T)
  ± 0.05%T at 220nm & 340nm
  ≤ 0.002A/hr at 500nm, 0A
  USB
  
85-240 VAC, 50/60 Hz
  
570 (L) x 460 (W) x 230 (H) mm
  14.5 kg